Chromosorb 101 HP60~80目填充柱 測定醫(yī)用口罩中痕量環(huán)氧乙烷GB 15979-2002
產(chǎn)品名稱:Chromosorb 101 HP60~80目填充柱 測定醫(yī)用口罩中痕量環(huán)氧乙烷GB 15979-2002
產(chǎn)品型號(hào):Chromosorb 101 HP60~80目填充柱
產(chǎn)品廠商:浩瀚色譜(山東)應(yīng)用技術(shù)開發(fā)有限公司
簡單介紹
Chromosorb 101 HP60~80目填充柱 測定醫(yī)用口罩中痕量環(huán)氧乙烷GB 15979-2002的詳細(xì)介紹
測定醫(yī)用口罩中痕量環(huán)氧乙烷GB 15979-2002 詳細(xì)信息:
柱:Chromosorb 101 HP60~80目;玻璃柱長2m,φ 3 mm。柱溫:120℃。
檢測器:150℃。
氣化器:150℃。
載氣量:氮?dú)猓?5 mL/min。
氫氣:35 mL/min。
空氣:350 mL/min。
柱前壓約為108 kPa。
應(yīng)用:GB 15979-2002 一次性使用衛(wèi)生用品衛(wèi)生標(biāo)準(zhǔn)
浩瀚色譜(山東)應(yīng)用技術(shù)開發(fā)有限公司建立了吹掃捕集-氣相色譜/質(zhì)譜法測定醫(yī)用口罩中痕量環(huán)氧乙烷的方法。采用吹掃捕集法對(duì)醫(yī)用口罩中環(huán)氧乙烷進(jìn)行富集,熱脫附后導(dǎo)入氣相色譜/質(zhì)譜儀,并選用選擇離子模式(SIM)進(jìn)行檢測,內(nèi)標(biāo)法定量。結(jié)果表明:環(huán)氧乙烷在5.0~200μg/L質(zhì)量濃度范圍內(nèi)線性關(guān)系良好,相關(guān)系數(shù)為0.9992,方法檢出限為0.03μg/g,定量限為0.10μg/g。三個(gè)不同濃度加標(biāo)水平(5.0μg/L、20μg/L、80μg/L)的回收率為97.59%~115.95%,相對(duì)標(biāo)準(zhǔn)偏差(RSD,n=6)為3.07%~7.48%。該方法操作簡單、富集效率高、快速準(zhǔn)確,適用于大批量醫(yī)用口罩樣品測定。
測定醫(yī)用口罩中痕量環(huán)氧乙烷GB 15979-2002 測試譜圖:
Determination of trace amounts of ethylene oxide in medical masks GB 15979-2002
Determination of trace amounts of ethylene oxide in medical masks GB 15979-2002 Details:
Apparatus: GC- gas chromatograph, hydrogen flame detector (FID).
Column: Chromosorb 101 HP60-80 mesh; glass column length 2m, φ 3 mm. Column temperature: 120°C.
Detector: 150°C.
Vaporizer: 150°C.
Carrier gas volume: nitrogen: 35 mL/min.
Hydrogen: 35 mL/min.
Air: 350 mL/min.
The pre-column pressure is a
Application: GB 15979-2002 Hygiene Standard for Disposable Hygiene Products
Haohan Chromatography (Shandong) Application Technology Development Co., Ltd. has established a method for the determination of trace amounts of ethylene oxide in medical masks by purge and trap-gas chromatography/mass spectrometry. The ethylene oxide in the medical masks was enriched by the purge and trap method, and then introduced into a gas chromatography/mass spectrometer after thermal desorption. The results showed that the linear relatio
Determination of trace ethylene oxide in medical masks GB 15979-2002 test spectrum: